Αρχειοθήκη ιστολογίου

Αλέξανδρος Γ. Σφακιανάκης
ΩτοΡινοΛαρυγγολόγος
Αναπαύσεως 5
Άγιος Νικόλαος Κρήτη 72100
2841026182
6032607174

Πέμπτη 13 Αυγούστου 2015

! ORL via Alexandros G.Sfakianakis on Inoreader: Evaluation of the rigidity of sagittal split ramus osteotomy fixation using four designs of biodegradable and titanium plates—a numerical study

! ORL via Alexandros G.Sfakianakis on Inoreader
 
Evaluation of the rigidity of sagittal split ramus osteotomy fixation using four designs of biodegradable and titanium plates—a numerical study
Aug 14th 2015, 00:20

Abstract

Purpose

This study was conducted to determine the best design of biodegradable plates for providing rigidity when used for fixation of sagittal split ramus osteotomy.

Methods

A computerized tomography image of a patient was used to generate a 3D model of a hemi-mandible. Four plate designs were merged with the hemi-mandible. They were (1) straight plate, (2) double straight plate, (3) T-shaped plate, and (4) double Y-shaped plate. Four finite element models were analyzed using the properties of biodegradable materials for the plates, and four additional models were analyzed using titanium alloy properties.

Results

The models predicted that rigidity of fixation would be noticeably less among biodegradable plates than titanium plates. They also predicted that the most rigid design among the titanium plates would be the straight plate, but among the biodegradable plates, it would be the double Y-shaped plate.

Conclusion

The double Y-shaped design is recommended when using biodegradable plates in fixation of sagittal split ramus osteotomy.