Αρχειοθήκη ιστολογίου

Αλέξανδρος Γ. Σφακιανάκης
ΩτοΡινοΛαρυγγολόγος
Αναπαύσεως 5
Άγιος Νικόλαος Κρήτη 72100
2841026182
6032607174

Πέμπτη 4 Μαρτίου 2021

Evaluation of image quality and assessability of a new flat‐panel 3D C‐arm compared to mobile and fixed computed tomography in posterior spinal fixation

xlomafota13 shared this article with you from Inoreader

Abstract

Background

This study evaluates image quality of a new flat‐panel three‐dimensional (3D) C‐arm to the current generation and to intraoperative and stationary computed tomography (CT) in spinal surgery.

Methods

Three‐dimensional‐imaging of 44 posterior screws acquired with two generations of 3D C‐arms (Arcadis Orbic, AO and Cios Spin, CS, Siemens, Germany) and CT scans (mobile intraoperative CT Airo, iCT, Brainlab, Germany and stationary CT Aquilion 32, sCT, Toshiba, Japan) were performed. Evaluation regarded assessability and measurements of implant position.

Results

Assessability score was 0.11 in AO, 0.56 in sCT, 0.91 in iCT and 1.46 in CS (p < 0.0005). AO and CS showed a significant difference in accuracy (p = 0.001) as well as CS and iCT (p < 0.001). Measurements of protrusion over the anterior edge did not show a significant difference (p = 0.341).

Conclusions

Image quality of the new generation of flat‐panel 3D C‐arms competes with CT imaging and offers significant advantages compared to the former generation.

View on the web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου