Lessons Learned from Implantation of Morcher 50D and 96S Artificial Iris Diaphragms

Purpose: To discuss problems associated with the implantation of two Morcher iris diaphragm models. Methods: We describe the history, intraoperative complications, and postoperative complications of 5 patients with specific Morcher iris implants. Results: We implanted Morcher 50D devices in 1 patient and Morcher 96S devices in 4 patients. Complications included postoperative rotation, device mis-sizing, difficult intraoperative rotation, zonular dehiscence, and intraoperative hemorrhage. Conclusion: Artificial iris implantation has a steep learning curve. With widespread availability on the horizon in the United States, the sharing of surgical experiences is key to achieving the best outcomes for patients.
Case Rep Ophthalmol 2017;8:527–534